FMA1600- SERIES MASS VOLUMETRIC FLOW METERS
The FMA-1600A Series mass and volumetric Omega’s flow meters use the principle of differential pressure within a laminar flow field to determine the mass flow rate. A laminar flow element (LFD) inside the meter forces the gas into laminar (streamlined) flow. Inside this region, the Poiseuille equation dictates that the volumetric flow rate be linearly related to the pressure drop along a fixed distance of the LFE. This, along with the viscosity of the gas, is used to accurately determine the volumetric flow rate. Separate absolute temperature and pressure sensors are incorporated and correct the volumetric flow rate to a set of standard conditions. This standardized flow rate is commonly called the mass flow rate and is reported in units such as standard cubic feet per minute (SCFM) or standard liters per minute (SLM).
Standard units include a 0 to 5 V output (4 to 20 mA optional) and RS232 communications. The gas select feature can be adjusted from the front of the keypad or via RS232 communications. Volumetric flow, mass flow, or absolute pressure, and temperature can be viewed or recorded through the RS232 connection. It is also possible to multi-drop up to 26 units on the same serial connection to a distance of 125′. These flowmeters are also available in a portable version for use in the field (-B models).
Specifications
FMA1600- SERIES MASS VOLUMETRIC FLOW METERS
- Accuracy± 0.8 %
- Accuracy Detail± (0.8% of rdg + 0.2% FS)
- Additional Features 1Power Supply Included
- Calibration NIST 5-Point Certificate
- Data InterfaceRS232
- Display Type LCD
- Electrical Connection8-pin mini DIN
- Height4.1 in
- Humidity Range0 to 100% RH non-condensing
- Materials, Wetted Aluminum, Brass, Epoxy, FKM, Glass, Gold, Silicone, Stainless Steel
- Measurable Flow Rate125%
- Media Compatibility Air, Gases, Nitrogen, Oxygen
- Operating Pressure, Max145 psi
- Operating Pressure, Min0 psi
- Operating Temperature, Max50 °C
- Operating Temperature, Min-10 °C
- Pressure Drop, Min1 psi
- Repeatability± 0.2%
- Response Time10 ms
- Response Time Note10 ms typical default response time for 63.2% of a step change. A variable register allows response time to be field adjustable to a certain extent via RS232 communications. The primary trade-off for response time is signal noise
- Span Shift0.02% FS/°C/atm
- Supply Current35 mA
- Supply Power Note35 mA typical current draw; 100 mA available supply recommended
- Turndown Ratio200:1
- Volumetric Flow Rate, Min0.05 SLM
- Width1.1 in
Accuracy: ± (0.8% of rdg + 0.2% FS)
Repeatability: ± 0.2%
Turndown Ratio: 200:1
Response Time: 10 ms typical default response time for 63.2% of a step change. A variable register allows response time to be field adjustable to a certain extent via RS232 communications. The primary trade-off for response time is signal noise
Output: 0 to 5 Vdc standard
Operating Temperature: -10 to 50°C (14 to 122°F)
Zero Shift: 0.02% FS/°C/atm
Span Shift: 0.02% FS/°C/atm
Humidity Range: 0 to 100% non-condensing
Pressure (Max):
FMA-1600A Series: 145 psig
FMA-1600A Series: 50 psi differential
FVL-1600A Series: 10 psig
FVL-1600A Series: 10 psi differential
Measurable Flow Rate: 125% FS
Supply Voltage: 7 to 30 Vdc (15 to 30 Vdc for 4 to 20 mA output)
Supply Current: 35 mA typical current draw; 100 mA available supply recommended
Cable Connection: 8-pin mini DIN
Wetted Parts: 303 & 302 Stainless Steel, FKM Heat Cured Silicone Rubber, Glass Reinforced Polyphenylene Sulfide, Heat Cured Epoxy, Aluminum, Gold, Silicon, Glass. Contact Omega engineering for details.
-LSS (Option):316LSS, 303SS, 430FRSS, Perfluoroelastomer standard, FKM, EPDM, Buna, Neoprene Standard Pressure Drops (venting to atmosphere):
| Max Flow Rate |
0.5 to 50 SCCM |
0.1 to 10 SLM |
20 SLM |
50 SLM |
100 SLM |
250 SLM |
500 SLM |
1000 SLM |
1500 SLM |
| PSID |
1 |
1 |
1 |
2 |
2.5 |
4.0 |
5.5 |
6.0 |
9.0 |